The AM-5100 SO2 Analyzer, based on the Ultraviolet Fluorescence Principle and combined with advanced Micro-processing Technology, provides accurate and reliable low-concentration SO2 detection of 'PPB~PPM' level, which can be applied to atmospheric monitoring and CEMS system (dilution method).
Features
Measurement range from 0~500ppb to 0~20ppm, selectable.
High efficiency hydrocarbon remover and special optical design, strong anti-interference ability.
Timed dark current compensation, high stability.
With self-check function of light intensity attenuation.
Adaptive signal filtering technology optimizes response time.
Large capacity memory, store historical data automatically.
Application
Applied to atmospheric monitoring.
Specifications
Measuring range: 0~500ppb, 0~20,000ppb (adjustable).
Unit: ppb, ppm, ug/m3, mg/m3
Zero noise:<0.2ppb(RMS)
Limit of detection: 0.4ppb
Zero drift:<1ppb/24h
Span drift:<5ppb/24h
Response time:<120s (T90)
Linearity:<1%/F.S.
Repeatability:<1%.
Sampling flow range: 800SCCM±10%.
Ambient temperature: 5℃~40℃.
Dimension: 178mmx430mmx635mm.
Weight:<15kg.
Power: 200V~240V, 50/60Hz, 400W.
Analog output: 10V, 5V, 1V, 0.1V (selectable).
Communication: RS232, RS485.